What does pacemaker capture mean?
A pacemaker consists of a box (i.e. the pacemaker or pulse generator) and a lead or leads. When it malfunctions, the issue is with rate, pacing, capturing (i.e. delivering the spike to depolarize the myocardium), or sensing [1].
What is capture on pacing?
Electrical capture occurs when a pacing stimulus leads to depolarization of the ventricles. It is confirmed by ECG changes typical of ventricular complexes — a widening of the QRS complex and a tall, broad T wave, — displayed on the monitor (See Figures 1–3).
What is failure to capture in a pacemaker?
Failure to capture occurs when a pacing stimulus is generated, but fails to trigger myocardial depolarization. On the ECG, failure to capture is identified by the presence of pacing spikes without associated myocardial depolarization.
What is pacing capture threshold?
The capture threshold is the minimum current setting required to produce a depolarisation of the paced chamber. In order to find this minimum current setting one can perform a simple manoeuvre: Set the pacemaker well above the native rate, so that the chamber of interest is being paced continuously.
How do you identify a pacemaker rhythm?
The pacemaker rhythm can easily be recognized on the ECG. It shows pacemaker spikes: vertical signals that represent the electrical activity of the pacemaker. Usually these spikes are more visible in unipolar than in bipolar pacing.
What is Oversensing?
Oversensing occurs when the pacemaker senses electrical signals that it should not normally encounter, which results in inappropriate inhibition of the pacing stimulus.
What rhythms can you pace?
Transcutaneous Pacing
- bradycardia unresponsive to drug therapy.
- 3rd degree heart block.
- Mobitz type II second-degree heart block when haemodynamically unstable or operation planned.
- overdrive pacing.
- asystole.
What are symptoms of pacemaker failure?
Clinical symptoms of pacemaker malfunction are variable and include syncope, dizziness, palpitations, and slow or fast heart rate. Extracardiac stimulation or hiccough may be present. Obtain as much information as possible regarding the pulse generator, leads, and programmed values.
How do you fix failure to pace?
Failure to pace• If the pulse generator is turned on but the indicators aren’t flashing, change the battery. If that doesn’t help, use a different pulse generator. If the pacing or indicator light flashes, check the connections to the cable and the position of the pacing electrode in the patient (done by X-ray).
How is capture associated with noninvasive pacing?
Capture accompanying noninvasive pacing resembles that seen in permanent or temporary invasive pacing. Mechanical capture is the contraction of the myocardium and is evidenced by presence of a pulse and signs of improved cardiac output. Both electrical and mechanical capture must occur to benefit the patient. Many patients achieve capture at 50 to
When does a pacing stimulus lead to capture?
Electrical capture occurs when a pacing stimulus leads to depolarization of the ventricles. It is confirmed by ECG changes typical of ventricular complexes — a widening of the QRS complex and a tall, broad T wave, — displayed on the monitor (See Figures 1–3). The deflection of the captured complex may be positive or negative. Capture
What is the definition of capture in medical dictionary?
Also found in: Dictionary, Thesaurus, Legal, Financial, Acronyms, Idioms, Encyclopedia, Wikipedia. capture. 1. the production of a ventricular complex from a supraventricular source following a period of atrioventricular dissociation. 2. in cardiac pacing terminology, the successful pacing of the heart by a pulse generator.
When does capture occur in a cardiac pacemaker?
capture Cardiac pacing Depolarization of the atria and/or ventricles by an electrical stimulus delivered by an artificial pacemaker; one-to-one capture occurs when each electrical stimulus causes a corresponding depolarization.